SPECIFICATOINS:
Infusion:
Overall Accuracy*: +/- 2% (Measured time > 1hr, min 0.2 ml)
Syringe Types:
|
BD
|
10/20/50/60 ml
|
|
Dispovan
|
10/20/50 ml
|
|
Terumo
|
10/30/50/60 ml
|
|
Pharma
|
10/50 ml
|
|
Braun Prefuser
|
50 ml
|
(Other brands are available on special request)
Flow Rate Ranges:
0.1 – 200 ml/hr for 50/60 ml syringe
0.1 – 120 ml/hr for 30 ml syringe
0.1 – 100 ml/hr for 20 ml syringe
0.1 – 060 ml/hr for 10 ml syringe
|
Infusion Volume limits:
|
60 ml, Programmable every 0.1 ml
|
|
Bolus Rate:
|
Up to 1000 ml/hr
|
|
KVO rate:
|
1 % of selected infusions rate but min. 0.1ml/hr
|
Occlusion pressure limit: Programmable three levels
|
Low
|
0.4 +/0.08 Kg/cm2 (bar)
|
|
Medium
|
0.9 +/0.20 Kg/cm2 (bar)
|
|
High
|
1.2 +/0.25 Kg/cm2 (bar)
|
Mechanical:
Driving Mechanism: Lead screw mechanism
Audio Visual Alarms:
|
Audio Visual Alarms:
|
No Mains, Very low battery, Low battery, Occlusion, No Syringe, Idle, Near End, End of infusion, KVO
|
|
Indicators:
|
Run & Mains
|
Special Features:
Program Save Option
You can save the infusion parameter which can recall by single key while new infusion saves the set up time
Run Time Change
All parameters can change during infusion without interrupting the infusion
Power Supply
|
External power supply:
|
AC Input range 170~ 270 V AC, Frequency 50 ~ 60 Hz, 12 Volts, 1 Amp DC output Supply
|
|
Power Consumption:
|
12 Watt max.
|
Classification:
Operating conditions:
|
Ambient Temperature
|
05 to 40° C
|
|
Relative Humidity
|
20 to 90 % (no condensation)
|
|
Atmospheric Pressure
|
70 to 106 kPa
|
Housing & Construction:
|
Housing Material:
|
Compact and Robust FRP
|
|
Weight:
|
Approx. 2.2 Kgs
|
|
Dimensions:
|
Approx. 265(L) mm, 110(H) mm, 150(W) mm
|
|
Mounting:
|
On a pole or a bed-frame
|
Note: Specifications and appearance of PLENUM SP-160D SYRINGE INFUSION PUMP are subjected to change without prior notice.