Welcome to Our Plenum Technologies
|
 |
|
Plenum's Volumetric Infusion Pump - Imac 304D
Infusion:
|
Overall Accuracy:
|
+/- 5% of the selected Infusion Rate with assigned I.V. set
|
|
Setting Modes:
|
Flow rate only, flow rate + volume target, time + volume target
|
|
Flow Rate Ranges:
|
0.1 – 999 ml/hr
|
|
Infusion Volume limits
|
0.1 - 9999 ml
|
|
Time range
|
0-99:59 hrs
|
|
Bolus Rate
|
100 – 1000 ml/hr adjustable in the steps of 100 ml/hr
|
|
Bolus Volume
|
0.1 – 9999 ml
|
|
Purge Rate
|
|
|
KVO rate
|
1 to 5 % of selected infusions rate but minimum ml/hr
|
Mechanical:
|
Pumping Mechanism:
|
Linear Peristaltic
|
Other:
|
Drop Sensor
|
External (Optional for the detection of upstream occlusion)
|
|
Air Bubble Detector
|
Ultrasonic, Adjustable bubble size from 50-150ml
Accumulated bubble size of 1ml over period of 15 minutes
|
|
Set Type
|
JMS, BBRAUN, BAXTER, POLIMED, ROMSON
Facility to configure two customizable sets as per user’s choice
|
|
Occlusion
|
Programmable 3 Levels
LOW 0.5 +/0.10 Kg/cm2 (bar)
MID 0.9 +/0.20 Kg/cm2 (bar)
HIGH 1.2 +/0.25 Kg/cm2 (bar)
|
Indicators & Alarms:
|
Audio Visual Alarms
|
No Mains, Very low battery, Low battery, Sensor error, Flow error, Door open, Near End, End of infusion, KVO, Air in line, Downstream Occlusion, Upstream Occlusion, Empty Container, System error, Pump stopped
|
|
Indicators
|
Run & Mains
|
Special Features:
|
Free Flow Protection
|
A unique lever mechanism protects from free flow in case of accidental door opening
|
|
Upstream Occlusion Detection (Optional)
|
Optional Drop sensor for the detection of upstream occlusion besides the downstream occlusion
|
|
Adjustable air in line Alarm threshold
|
The size of air bubble to be detected can be set in the range of 50 to 150ml
|
|
Bolus and Purge facility
|
Adjustable bolus rate and bolus volume alongwith the Purge facility
|
|
Keypad locking
|
A built in safely feature prevents any change the parameter if any key is accidentally pressed during infusion
|
Power Supply:
|
External power supply
|
AC Input range 170~ 270 V AC, Frequency 50 ~ 60 Hz, 12 Volts, 3 Amp DC output Supply
|
|
Power Consumption
|
30 Watt max.
|
|
Build in battery
|
8.4 V, 2.5 Ahr Ni-MH battery for increased durability
|
|
Battery life
|
Continuous operation for about 3.5 hrs. at an operating rate of 25 ml/hr with assign I.V. Set under standard conditions
|
Classification:
Operating conditions:
|
Ambient Temperature
|
05 to 40° C
|
|
Relative Humidity
|
20 to 90 % (no condensation)
|
|
Atmospheric Pressure
|
70 to 106 kPa
|
Housing & Construction
|
Housing Material
|
Compact and Robust ABS Plastic
|
|
Weight
|
Approx. 2.4 Kgs
|
|
Dimensions
|
Approx. 250(L) mm, 145(H) mm, 90(W) mm
|
|
Mounting
|
On a pole or a bed-frame
|
Note: Specifications and appearance are subjected to change without prior notice.
|
|
|
 |
|
 |
|
|
|
|